Characterization of spherically focused transducers using an ultrasonic measurement model approach

T. P. Lerch, L. W. Schmerr, A. Sedov

Research output: Contribution to journalArticlepeer-review

14 Scopus citations

Abstract

A new and computationally efficient method is developed for characterizing a spherically focused, ultrasonic transducer (and its accompanying test system). Procedures for determining the probe’s effective radius, effective focal length, and system efficiency factor are described. Predicted responses that make use of these effective parameters are shown to correspond very wellto measured responses for a number of different transducers.