TY - JOUR
T1 - Comparative Chronic Valve and Venous Effects of Lumenless versus Stylet-Delivered Pacing Leads in Patients with and Without Congenital Heart
AU - Bharmanee, Apinya
AU - Zelin, Kathleen
AU - Sanil, Yamuna
AU - Gupta, Pooja
AU - Karpawich, Peter P.
N1 - Publisher Copyright:
© 2015 Wiley Periodicals, Inc.
PY - 2015/11
Y1 - 2015/11
N2 - Background Standard, 5-7-Fr diameter pacing leads (PLs) can adversely affect atrioventricular valve (AVV) and venous (superior vena cava [SVC], innominate [INN]) integrities. Although chronic pacing/sensing performances have been reported on the steroid-eluting, lumenless, 4.1-Fr PL (Model 3830, Medtronic Inc., Minneapolis, MN, USA), comparative valve and venous effects are largely unknown. Methods Patients (n = 134) were divided into two PL groups: Group 1 (n = 65, Model 3830) and Group 2 (n = 69, various 5-7 Fr models) and followed up to 9 years postimplant. Patient demographics, clinical findings, valve function, and venous dimensions were reviewed. Statistical significance was defined as P < 0.05. Results Patient implant age (mean 16.4 years vs 17.3 years), presence of congenital heart defect (CHD), and preexisting valve issues were comparable between groups. New or worsening valve insufficiency occurred in 12% of Group 1 patients (mean follow-up 4.3 ± 2.8 years) and 27% of Group 2 patients (mean follow-up 6.2 ± 3.5 years; P < 0.05). Significant SVC or INN narrowing was found in 11 % of Group 1 and 24% of Group 2 patients (P = 0.0004). All Group 1 patients <12 years of age showed normal while 50% of those from Group 2 exhibited stunted SVC or INN growth (P < 0.05). Conclusion The lumenless, 4.1-Fr diameter PL offers improved clinical benefits, better AVV integrity, and venous development compared with larger 5-7-Fr diameter PL and should be considered especially in younger patients with/without CHD.
AB - Background Standard, 5-7-Fr diameter pacing leads (PLs) can adversely affect atrioventricular valve (AVV) and venous (superior vena cava [SVC], innominate [INN]) integrities. Although chronic pacing/sensing performances have been reported on the steroid-eluting, lumenless, 4.1-Fr PL (Model 3830, Medtronic Inc., Minneapolis, MN, USA), comparative valve and venous effects are largely unknown. Methods Patients (n = 134) were divided into two PL groups: Group 1 (n = 65, Model 3830) and Group 2 (n = 69, various 5-7 Fr models) and followed up to 9 years postimplant. Patient demographics, clinical findings, valve function, and venous dimensions were reviewed. Statistical significance was defined as P < 0.05. Results Patient implant age (mean 16.4 years vs 17.3 years), presence of congenital heart defect (CHD), and preexisting valve issues were comparable between groups. New or worsening valve insufficiency occurred in 12% of Group 1 patients (mean follow-up 4.3 ± 2.8 years) and 27% of Group 2 patients (mean follow-up 6.2 ± 3.5 years; P < 0.05). Significant SVC or INN narrowing was found in 11 % of Group 1 and 24% of Group 2 patients (P = 0.0004). All Group 1 patients <12 years of age showed normal while 50% of those from Group 2 exhibited stunted SVC or INN growth (P < 0.05). Conclusion The lumenless, 4.1-Fr diameter PL offers improved clinical benefits, better AVV integrity, and venous development compared with larger 5-7-Fr diameter PL and should be considered especially in younger patients with/without CHD.
KW - clinical
KW - echocardiography
KW - electrophysiology
KW - pacing
KW - pediatrics
UR - http://www.scopus.com/inward/record.url?scp=84945313661&partnerID=8YFLogxK
U2 - 10.1111/pace.12728
DO - 10.1111/pace.12728
M3 - Article
C2 - 26256093
AN - SCOPUS:84945313661
VL - 38
SP - 1343
EP - 1350
JO - Pacing and Clinical Electrophysiology
JF - Pacing and Clinical Electrophysiology
SN - 0147-8389
IS - 11
ER -