TY - JOUR
T1 - Development of an active ankle foot orthosis to prevent foot drop and toe drag in hemiplegic patients
T2 - A preliminary study
AU - Kim, Jungyoon
AU - Hwang, Sungjae
AU - Sohn, Ryanghee
AU - Lee, Younghee
AU - Kim, Youngho
PY - 2011
Y1 - 2011
N2 - We developed an active ankle-foot orthosis (AAFO) that controls dorsiflexion/plantarflexion of the ankle joint to prevent foot drop and toe drag during hemiplegic walking. To prevent foot slap after initial contact, the ankle joint must remain active to minimize forefoot collision against the ground. During late stance, the ankle joint must also remain active to provide toe clearance and to aid with push-off. We implemented a series elastic actuator in our AAFO to induce ankle dorsiflexion/plantarflexion. The activator was controlled by signals from force sensing register (FSR) sensors that detected gait events. Three dimensional gait analyses were performed for three hemiplegic patients under three different gait conditions: gait without AFO (nAFO), gait with a conventional hinged AFO that did not control the ankle joint (HAFO), and gait with the newly-developed AFO (AAFO). Our results demonstrate that our newly-developed AAFO not only prevents foot drop by inducing plantarflexion during loading response, but also prevents toe drag by facilitating plantarflexion during pre-swing and dorsiflexion during swing phase, leading to improvement in most temporal-spatial parameters. However, only three hemiplegic patients were included in this gait analysis. Studies including more subjects will be required to evaluate the functionality of our newly developed AAFO.
AB - We developed an active ankle-foot orthosis (AAFO) that controls dorsiflexion/plantarflexion of the ankle joint to prevent foot drop and toe drag during hemiplegic walking. To prevent foot slap after initial contact, the ankle joint must remain active to minimize forefoot collision against the ground. During late stance, the ankle joint must also remain active to provide toe clearance and to aid with push-off. We implemented a series elastic actuator in our AAFO to induce ankle dorsiflexion/plantarflexion. The activator was controlled by signals from force sensing register (FSR) sensors that detected gait events. Three dimensional gait analyses were performed for three hemiplegic patients under three different gait conditions: gait without AFO (nAFO), gait with a conventional hinged AFO that did not control the ankle joint (HAFO), and gait with the newly-developed AFO (AAFO). Our results demonstrate that our newly-developed AAFO not only prevents foot drop by inducing plantarflexion during loading response, but also prevents toe drag by facilitating plantarflexion during pre-swing and dorsiflexion during swing phase, leading to improvement in most temporal-spatial parameters. However, only three hemiplegic patients were included in this gait analysis. Studies including more subjects will be required to evaluate the functionality of our newly developed AAFO.
KW - Active ankle-foot-orthosis
KW - dorsiflexion
KW - foot drop
KW - plantarflexion
KW - toe drag
UR - http://www.scopus.com/inward/record.url?scp=80155147242&partnerID=8YFLogxK
U2 - 10.3233/ABB-2011-0008
DO - 10.3233/ABB-2011-0008
M3 - Article
AN - SCOPUS:80155147242
SN - 1176-2322
VL - 8
SP - 377
EP - 384
JO - Applied Bionics and Biomechanics
JF - Applied Bionics and Biomechanics
IS - 3-4